M-001 Dual Channel Serial Injection - Analyst with ICC-CE

From LEAP

(Difference between revisions)
Jump to: navigation, search
Current revision (20:19, 28 January 2010) (view source)
(For other cost/time saving measures consider:)
 
(5 intermediate revisions not shown.)
Line 8: Line 8:
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]]
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]]
=== Overview ===
=== Overview ===
-
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within '''Analyst®  Software'''[https://products.appliedbiosystems.com/ab/en/US/adirect/ab?cmd=catNavigate2&catID=600927&tab=DetailInfo] from ''Applied Biosystems Inc® '' using a LEAP Cycle (requires ICC-CE software) [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-software/instrument-control-components-and-cycle-editor-181.php].
+
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within [https://products.appliedbiosystems.com/ab/en/US/adirect/ab?cmd=catNavigate2&catID=600927&tab=DetailInfo '''Analyst®  Software'''] from ''Applied Biosystems Inc® '' using a LEAP Cycle [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-software/instrument-control-components-and-cycle-editor-181.php (requires ICC-CE software)].
-
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration [http://leapwiki.com/mediawiki/index.php?title=Category:Regeneration]. This configuration requires a 6 port valve[http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/6-port-2-position-valve-155.php] for sample loading and a 10 port valve [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/10-port-2-position-valve-157.php]for column switching. Both valves must be controlled by the LEAP cycle.   
+
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for [http://leapwiki.com/mediawiki/index.php?title=Category:Regeneration regeneration]. This configuration requires a 6 port [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/6-port-2-position-valve-155.php valve] for sample loading and a 10 port [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/10-port-2-position-valve-157.php valve]for column switching. Both valves must be controlled by the LEAP cycle.   
Line 24: Line 24:
{|  
{|  
|-
|-
-
| Saving of Reagent time (Back cut)[http://www.leapwiki.com/mediawiki/index.php?title=Category:Backcut]
+
| Saving of Reagent time [http://www.leapwiki.com/mediawiki/index.php?title=Category:Backcut (Back cut)]
|-
|-
| Increase throughput at little cost
| Increase throughput at little cost
Line 40: Line 40:
{|  
{|  
|-
|-
-
| Cannot do "Look Ahead"[http://www.leapwiki.com/mediawiki/index.php?title=Category:Look_Ahead]
+
| Cannot do [http://www.leapwiki.com/mediawiki/index.php?title=Category:Look_Ahead "Look Ahead"]
|-
|-
-
| Injections are not staggered[http://www.leapwiki.com/mediawiki/index.php?title=Category:Staggered_Injections]
+
| Injections are not [http://www.leapwiki.com/mediawiki/index.php?title=Category:Staggered_Injections staggered]
|-
|-
| Peaks of interest must come off early
| Peaks of interest must come off early
Line 86: Line 86:
=== For other cost/time saving measures consider:===
=== For other cost/time saving measures consider:===
-
* LEAP offers multiple choices of Valve Self Wash Stations to aid in cleaning the valve [http://www.leapwiki.com/mediawiki/index.php?title=Valve_Washing]
+
* [http://www.leapwiki.com/mediawiki/index.php?title=Valve_Washing LEAP offers multiple choices of Valve Self Wash Stations to aid in cleaning the valve]
-
* Dual Channel Serial Injection with Front Cut (M-002) using Analyst and ICC-CE [http://www.leapwiki.com/mediawiki/index.php?title=M-002_Dual_Channel_Serial_Injection_with_front_cut_-_Analyst_with_ICC-CE]
+
* [http://www.leapwiki.com/mediawiki/index.php?title=M-002_Dual_Channel_Serial_Injection_with_front_cut_-_Analyst_with_ICC-CE Dual Channel Serial Injection with Front Cut (M-002) using Analyst and ICC-CE ]
-
* Dual Channel Parallel Injection (M-003) using Cycle Composer Software and Analyst [http://www.leapwiki.com/mediawiki/index.php?title=M-003_Dual_Channel_Parallel_Injection_-_Cycle_Composer]
+
* [http://www.leapwiki.com/mediawiki/index.php?title=M-003_Dual_Channel_Parallel_Injection_-_Cycle_Composer Dual Channel Parallel Injection (M-003) using Cycle Composer Software and Analyst]
-
* “Look Ahead”, sample prep before injection (M-004) using LEAP Shell Software [http://www.leaptec.com/software/leap-shell-3.php]
+
* [http://www.leaptec.com/software/leap-shell-3.php “Look Ahead”, sample prep before injection (M-004) using LEAP Shell Software]
-
* Dual Channel Parallel Injection (M-005) using LEAP Shell software [http://www.leapwiki.com/mediawiki/index.php?title=M-005_Dual_Channel_Parallel_Injection_-_LEAP_Shell_%22(M-Fab)%22]
+
* [http://www.leapwiki.com/mediawiki/index.php?title=M-005_Dual_Channel_Parallel_Injection_-_LEAP_Shell_%22(M-Fab)%22 Dual Channel Parallel Injection (M-005) using LEAP Shell software]
-
* Summary of LC Multi Valve Applications [http://www.leapwiki.com/mediawiki/index.php?title=Summary_of_LC_Multi_Valve_Applications]
+
* [http://www.leapwiki.com/mediawiki/index.php?title=Summary_of_LC_Multi_Valve_Applications Summary of LC Multi Valve Applications ]
-
* LEAP Shell Software and application for custom injections and scheduling [http://www.leaptec.com/software/leap-shell-3.php]
+
* [http://www.leaptec.com/software/leap-shell-3.php LEAP Shell Software and application for custom injections and scheduling ]
 +
 
 +
=== Accessories for the PAL ===
 +
[[image:Accessories_Icon.png|40px]] [http://www.leapwiki.com/mediawiki/index.php?title=Category:Accessories Other Accessories for PAL Robots]
 +
 
 +
 
----
----
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]]
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]]

Current revision

LC Column Switching & Regeneration
Application Type
  SPECIAL
Application ID
  M-001
Description
  LC Column Switching & Regeneration.

Contents

Overview

This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within Analyst® Software from Applied Biosystems Inc® using a LEAP Cycle (requires ICC-CE software).

There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valvefor column switching. Both valves must be controlled by the LEAP cycle.


Please note: LEAP can not control the MS Selector valve as part of the cycle.

Plumbing of valve diagram showing flow for each selected column with the regeneration of the analytical column while the other is engaged with the LCMS system
Plumbing of valve diagram showing flow for each selected column with the regeneration of the analytical column while the other is engaged with the LCMS system







Key Benefits
Saving of Reagent time (Back cut)
Increase throughput at little cost
Control is all within Analyst Software
Compliance not affected
GLP Compliant



Restrictions
Cannot do "Look Ahead"
Injections are not staggered
Peaks of interest must come off early
Columns may perform slightly different on alternating samples








Code Application Control SW Key Hardware Key Benefits Restrictions GLP
M-001 Dual Channel Serial Injection

Provides shorter cycle times by regenerating columns during next analytical run
Analyst with ICC-CE * 6 port injection valve
* 10 port Valve
* Gradient Pump
* Regeneration pump
* Saving of regeneration time(Back Cut)
* Increase throughput at little cost
* Control is all within Analyst
* Compliance is not affected
* Can not do "Look Ahead"
* Injections are not staggered
* Peaks of interest must come of early
* Columns may perform slightly differently on alternating samples
Yes



Typical System of PAL with CoolStack Fast Wash and Valves

Typical system for Dual Column with Regeneration


When columns must be regenerated following each analysis, this technique permits automation of the process. While one column performs the analysis, the second column undergoes regeneration by use of an auxiliary pump. Once the first analysis is complete, the valve is switched and the regenerated column is now ready for analytical use.

PRINT Function for Page

PDF Summary of New System

PDF Summary of Upgrade to Existing System


For other cost/time saving measures consider:

Accessories for the PAL

Other Accessories for PAL Robots



This workstation application is based on the CTC PAL X, Y, Z syringe only autosampler robot from LEAP Technologies. This extremely flexible, precise, and adaptable robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your LC or GC requirements. Please contact LEAP Technologies on how we can help you get maximized throughput with flexible automation.

Contact LEAP

Contact the LEAP office

Keywords: Dual Channel Serial Injection, Provides shorter cycle times by regenerating columns during next analytical run, Analyst with ICC-CE, 6 port injection valve, 10 port Valve, Gradient Pump ,Regeneration pump, Saving of regeneration time, Back Cut, Increase throughput at little cost, Control is all within Analyst, Compliance is not affected, Can not do Look Ahead, Injections are not staggered, Peaks of interest must come of early, Columns may perform slightly differently on alternating samples, ICC-CE software, cycles