M-001 Dual Channel Serial Injection - Analyst with ICC-CE
From LEAP
Line 8: | Line 8: | ||
[[Image:LEAP Logo.png|frame|www.leaptec.com<br>www.leapwiki.com]] | [[Image:LEAP Logo.png|frame|www.leaptec.com<br>www.leapwiki.com]] | ||
=== Overview === | === Overview === | ||
- | This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within '''Analyst® Software'''[https://products.appliedbiosystems.com/ab/en/US/adirect/ab?cmd=catNavigate2&catID=600927&tab=DetailInfo] from ''Applied Biosystems Inc® '' using a | + | This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within '''Analyst® Software'''[https://products.appliedbiosystems.com/ab/en/US/adirect/ab?cmd=catNavigate2&catID=600927&tab=DetailInfo] from ''Applied Biosystems Inc® '' using a LEAP Cycle (requires ICC-CE software) [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-software/instrument-control-components-and-cycle-editor-181.php]. |
- | There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve[http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/6-port-2-position-valve-155.php] for sample loading and a 10 port valve [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/10-port-2-position-valve-157.php]for column switching. Both valves must be controlled by the | + | There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve[http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/6-port-2-position-valve-155.php] for sample loading and a 10 port valve [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/10-port-2-position-valve-157.php]for column switching. Both valves must be controlled by the LEAP cycle. |
''Please note: LEAP can not control the MS Selector valve as part of the cycle.'' | ''Please note: LEAP can not control the MS Selector valve as part of the cycle.'' |
Revision as of 23:02, 28 April 2009
LC Column Switching & Regeneration | |
Application Type | |
SPECIAL | |
Application ID | |
M-001 | |
Description | |
LC Column Switching & Regeneration. |
Contents |
Overview
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within Analyst® Software[1] from Applied Biosystems Inc® using a LEAP Cycle (requires ICC-CE software) [2].
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve[3] for sample loading and a 10 port valve [4]for column switching. Both valves must be controlled by the LEAP cycle.
Please note: LEAP can not control the MS Selector valve as part of the cycle.
Key Benefits |
Saving of Reagent time (Back cut) |
Increase throughput at little cost |
Control is all within Analyst Software |
Compliance not affected |
GLP Compliant |
Restrictions |
Cannot do "Look Ahead" |
Injections are not staggered |
Peaks of interest must come off early |
Columns may perform slightly different on alternating samples |
Code | Application | Control SW | Key Hardware | Key Benefits | Restrictions | GLP |
---|---|---|---|---|---|---|
M-001 | Dual Channel Serial Injection Provides shorter cycle times by regenerating columns during next analytical run | Analyst with ICC-CE | * 6 port injection valve * 10 port Valve * Gradient Pump * Regeneration pump | * Saving of regeneration time(Back Cut) * Increase throughput at little cost * Control is all within Analyst * Compliance is not affected | * Can not do "Look Ahead" * Injections are not staggered * Peaks of interest must come of early * Columns may perform slightly differently on alternating samples | Yes |
PRINT Function for Page
For other cost/time saving measures consider:
- LEAP offers multiple choices of Valve Self Wash Stations to aid in cleaning the valve [5]
- Dual Channel Serial Injection with Front Cut (M-002) using Analyst and ICC-CE [6]
- Dual Channel Parallel Injection (M-003) using Cycle Composer Software and Analyst [7]
- “Look Ahead”, sample prep before injection (M-004) using LEAP Shell Software [8]
- Dual Channel Parallel Injection (M-005) using LEAP Shell software [9]
- LEAP Shell Software and application for custom injections and scheduling [10]
Contact LEAP
For additional information about this technique please contact LEAP Technologies for detailed information