M-001 Dual Channel Serial Injection - Analyst with ICC-CE

From LEAP

(Difference between revisions)
Jump to: navigation, search
Line 48: Line 48:
<br><br><br><br><br><br>
<br><br><br><br><br><br>
-
{| border="1"
+
 
-
|-valign="top"
+
{| cellpadding="6" style="border:1px solid darkgray;"
-
|'''Quick Overview'''
+
|- align=left
-
|-valign="top"
+
!width="25"|Code
-
! Code !! Application!!  Control SW !Key Hardware!Key Benefits !Restrictions !! GLP  
+
!width="200"|Application
-
|-valign="top"
+
!width="90"|Control SW
-
! M-001 !! Column switching with parallel regeneration <br>Provides shorter cycle times by regenerating columns during next analytical run!!  Analyst with ICC-CE !!  PAL<BR>6 port injection valve<BR>10 port Valve <br>Gradient Pump and Regeneration pump!!  Saving of regeneration time(Back Cut)<br> Increase throughput at little cost<br> Control is all within Analyst<br> Compliance is not affected !!  Can not do "Look Ahead"<br> Injections are not staggered<br> Peaks of interest must come of early<br>Columns may perform slightly differently on alternating samples!! Yes
+
!width="200"|Key Hardware
-
|-
+
!width="200"|Key Benefits
 +
!width="200"|Restrictions
 +
!width="20"|GLP  
 +
|- align=left
 +
| style="border:1px solid blue;"|'''M-001'''
 +
| style="border:1px solid blue;"|'''Column switching with parallel regeneration'''<br><br>Provides shorter cycle times by regenerating columns during next analytical run
 +
| style="border:1px solid blue;"|Analyst with ICC-CE  
 +
| style="border:1px solid blue;"|* 6 port injection valve<BR>* 10 port Valve <br>* Gradient Pump<br>* Regeneration pump
 +
| style="border:1px solid blue;"|* Saving of regeneration time(Back Cut)<br>* Increase throughput at little cost<br> * Control is all within Analyst<br>* Compliance is not affected
 +
| style="border:1px solid blue;"|* Can not do "Look Ahead"<br>* Injections are not staggered<br>* Peaks of interest must come of early<br>* Columns may perform slightly differently on alternating samples
 +
| style="border:1px solid blue;"|Yes
|}
|}
-
<br>
+
 
 +
 
=== PRINT Function for Page ===
=== PRINT Function for Page ===

Revision as of 21:00, 28 April 2009

LC Column Switching & Regeneration
Application Type
  SPECIAL
Application ID
  M-001
Description
  LC Column Switching & Regeneration.
www.leaptec.comwww.leapwiki.com
www.leaptec.com
www.leapwiki.com

Contents

Overview

This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within Analyst Software[1] from Applied Biosystems Inc using a Standard LEAP Cycle (requires ICC-CE software) [2].

There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve[3] for sample loading and a 10 port valve [4]for column switching. Both valves must be controlled by the Standard LEAP cycle. LEAP can not control the MS Selector valve as part of the cycle.

Plumbing of valve diagram showing flow for each selected column with the regeneration of the analytical column while the other is engaged with the LCMS system
Plumbing of valve diagram showing flow for each selected column with the regeneration of the analytical column while the other is engaged with the LCMS system





Key Benefits
Saving of Reagent time (Back cut)
Increase throughput at little cost
Control is all within Analyst Software
Compliance not affected
GLP Compliant



Restrictions
Cannot do "Look Ahead"
Injections are not staggered
Peaks of interest must come off early
Columns may perform slightly different on alternating samples








Code Application Control SW Key Hardware Key Benefits Restrictions GLP
M-001 Column switching with parallel regeneration

Provides shorter cycle times by regenerating columns during next analytical run
Analyst with ICC-CE * 6 port injection valve
* 10 port Valve
* Gradient Pump
* Regeneration pump
* Saving of regeneration time(Back Cut)
* Increase throughput at little cost
* Control is all within Analyst
* Compliance is not affected
* Can not do "Look Ahead"
* Injections are not staggered
* Peaks of interest must come of early
* Columns may perform slightly differently on alternating samples
Yes


PRINT Function for Page

For other cost/time saving measures consider:

  • LEAP offers multiple choices of Valve Self Wash Stations to aid in cleaning the valve [5]
  • Dual Channel Serial Injection with Front Cut (M-002) using Analyst and ICC-CE [6]
  • Dual Channel Parallel Injection (M-003) using Cycle Composer Software and Analyst [7]
  • “Look Ahead”, sample prep before injection (M-004) using LEAP Shell Software [8]
  • Dual Channel Parallel Injection (M-005) using LEAP Shell software [9]
  • LEAP Shell Software and application for custom injections and scheduling [10]

www.leaptec.comwww.leapwiki.com
www.leaptec.com
www.leapwiki.com

Contact LEAP

For additional information about this technique please contact LEAP Technologies for detailed information