M-001 Dual Channel Serial Injection - Analyst with ICC-CE
From LEAP
Line 10: | Line 10: | ||
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within '''Analyst Software'''[https://products.appliedbiosystems.com/ab/en/US/adirect/ab?cmd=catNavigate2&catID=600927&tab=DetailInfo] from ''Applied Biosystems Inc'' using a Standard LEAP Cycle (requires ICC-CE software) [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-software/instrument-control-components-and-cycle-editor-181.php]. | This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within '''Analyst Software'''[https://products.appliedbiosystems.com/ab/en/US/adirect/ab?cmd=catNavigate2&catID=600927&tab=DetailInfo] from ''Applied Biosystems Inc'' using a Standard LEAP Cycle (requires ICC-CE software) [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-software/instrument-control-components-and-cycle-editor-181.php]. | ||
- | There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching. Both valves must be controlled by the Standard LEAP cycle. LEAP can not control the MS Selector valve as part of the cycle. | + | There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve[http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/6-port-2-position-valve-155.php] for sample loading and a 10 port valve [http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-lc-pal-injection-valves/10-port-2-position-valve-157.php]for column switching. Both valves must be controlled by the Standard LEAP cycle. LEAP can not control the MS Selector valve as part of the cycle. |
[[Image:M-001.png|frame|Plumbing of valve diagram showing flow for each selected column with the regeneration of the analytical column while the other is engaged with the LCMS system]] | [[Image:M-001.png|frame|Plumbing of valve diagram showing flow for each selected column with the regeneration of the analytical column while the other is engaged with the LCMS system]] |
Revision as of 19:49, 28 April 2009
LC Column Switching & Regeneration | |
Application Type | |
SPECIAL | |
Application ID | |
M-001 | |
Description | |
LC Column Switching & Regeneration. |
Contents |
Overview
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within Analyst Software[1] from Applied Biosystems Inc using a Standard LEAP Cycle (requires ICC-CE software) [2].
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve[3] for sample loading and a 10 port valve [4]for column switching. Both valves must be controlled by the Standard LEAP cycle. LEAP can not control the MS Selector valve as part of the cycle.
Key Benefits |
Saving of Reagent time (Back cut) |
Increase throughput at little cost |
Control is all within Analyst Software |
Compliance not affected |
GLP Compliant |
Restrictions |
Cannot do "Look Ahead" |
Injections are not staggered |
Peaks of interest must come off early |
Columns may perform slightly different on alternating samples |
Quick Overview | ||||||
Code | Application | Control SW | Key Hardware | Key Benefits | Restrictions | GLP |
---|---|---|---|---|---|---|
M-001 | Column Switching with parallel regeneration Provides shorter cycle times by regenerating columns during next analytical run | Analyst with ICC-CE | PAL 10 port Valve Gradient Pump and Regeneration pump | Saving of regeneration time(Back Cut) | Can not do "Look Ahead" | Yes |
PRINT Function for Page
For other cost/time saving measures consider:
- LEAP offers multiple choices of Valve Self Wash Stations to aid in cleaning the valve [5]
- Dual Channel Serial Injection with Front Cut (M-002) using Analyst and ICC-CE [6]
- Dual Channel Parallel Injection (M-003) using Cycle Composer Software and Analyst [7]
- “Look Ahead”, sample prep before injection (M-004) using LEAP Shell Software [8]
- Dual Channel Parallel Injection (M-005) using LEAP Shell software [9]
- LEAP Shell Software and application for custom injections and scheduling [10]