Multi-Valve system for selectable loop size
From LEAP
(Difference between revisions)
Line 8: | Line 8: | ||
=== Contact LEAP === | === Contact LEAP === | ||
- | * [http://www.leaptec.com/ | + | * [http://www.leaptec.com/contact-us.php Contact the LEAP office] |
Revision as of 14:21, 21 April 2009
Two injection valves each plumbed with different loop size. Possible within Analyst to choose between a large or small injection loop by selecting different injection valves in the acquisition method.
Avoids having to re-plumb the valves for the different injection sizes. Two Analyst cycles were created, and a second six-port valve added to the PAL, also a 10-port serial/selector vial was added to dynamically make the selection between the 10ul and 100ul LC valves.
Project Ref: H1098'