M-005 Dual Channel Parallel Injection - LEAP Shell "(M-Fab)"
From LEAP
(→Hardware Requirements) |
|||
Line 66: | Line 66: | ||
| style="border:1px solid blue;"|'''Dual Channel Parallel Injection'''<br><br>Provides minimum cycle times by limiting the acquisition time to just the peaks of interest and regenerating columns during acquisition | | style="border:1px solid blue;"|'''Dual Channel Parallel Injection'''<br><br>Provides minimum cycle times by limiting the acquisition time to just the peaks of interest and regenerating columns during acquisition | ||
| style="border:1px solid blue;"|LEAP Shell | | style="border:1px solid blue;"|LEAP Shell | ||
- | | style="border:1px solid blue;"|*DoubleStack valve<br>*6 port injection valve<br>*FWS in front of 6 port valve<br>* Gradient Pump<br>* Regeneration pump | + | | style="border:1px solid blue;"|* DoubleStack valve<br>*6 port injection valve<br>* FWS in front of 6 port valve<br>* Gradient Pump<br>* Regeneration pump |
- | | style="border:1px solid blue;"|*Integration with Analyst Software<br>*Single Sample List<br> *Provides staggering of injections<br> *Multiplexes 2 separate LC systems into on MS | + | | style="border:1px solid blue;"|* Integration with Analyst Software<br>*Single Sample List<br> * Provides staggering of injections<br> * Multiplexes 2 separate LC systems into on MS |
- | | style="border:1px solid blue;"|* Does not support 21CFR11 compliance<br>* | + | | style="border:1px solid blue;"|* Does not support 21CFR11 compliance<br>* equires pumps to be integrated with LEAP Shell |
| style="border:1px solid blue;"|No | | style="border:1px solid blue;"|No | ||
|} | |} |
Revision as of 20:48, 11 May 2009
Dual Channel Parallel Injections with LEAP Shell "M-Fab" | |
Application Type | |
SPECIAL | |
Application ID | |
M-005 | |
Description | |
High Throughput LC injections alternating onto two analytical columns. |
Contents |
Overview
This configuration provides short cycle times by limiting the acquisition time to just the peaks of interest and regenerating the columns during the acquisition. It uses LEAP Shell Scheduling Software[1] - this application is integrated withAnalyst® Software[2] from Applied Biosystems Inc®. There is a single sample list and allows multiplexing [3] of 2 separate HPLC systems into one mass spectrometer.
If the peaks of interest come off together in the middle of the gradient (as shown below) then it is possible to utilize the front cut and back cut time to increase throughput.
If injections are staggered in the following way, there is a potential for up to 100% gain:
Hardware Requirements
- DoubleStack multiple valve system and a 6 port injection valve with FWS in front of it is required
Key Benefits |
Integration with Analyst |
Single Sample List |
Staggering of Samples [4] |
Multiplexes 2 separate LC systems into one Mass Spec |
Restrictions |
Does not yet support 21CFR11 |
Requires Pumps to be integrated with LEAP Shell |
Code | Application | Control SW | Key Hardware | Key Benefits | Restrictions | GLP |
---|---|---|---|---|---|---|
M-005 | Dual Channel Parallel Injection Provides minimum cycle times by limiting the acquisition time to just the peaks of interest and regenerating columns during acquisition | LEAP Shell | * DoubleStack valve *6 port injection valve * FWS in front of 6 port valve * Gradient Pump * Regeneration pump | * Integration with Analyst Software *Single Sample List * Provides staggering of injections * Multiplexes 2 separate LC systems into on MS | * Does not support 21CFR11 compliance * equires pumps to be integrated with LEAP Shell | No |
For other cost/time saving measures consider:
- LEAP offers multiple choices of Valve Self Wash Stations to aid in cleaning the valve [5]
- Dual Channel Serial Injection (M-001) using Analyst and ICC-CE [6]
- Dual Channel Serial Injection with Front Cut (M-002) using Analyst and ICC-CE [7]
- “Look Ahead”, sample prep before injection (M-004) using LEAP Shell Software [8]
- LEAP Shell Software and application for custom injections and scheduling [9]
Contact LEAP
For additional information about this technique please contact LEAP Technologies for detailed information