M-003 Dual Channel Parallel Injection - Cycle Composer
From LEAP
(Difference between revisions)
Line 10: | Line 10: | ||
=== Overview === | === Overview === | ||
This configuration allows shorter cycle times by limiting the acquisition time to just the peaks of interest and regenerating the alternating column during the acquisition. It uses | This configuration allows shorter cycle times by limiting the acquisition time to just the peaks of interest and regenerating the alternating column during the acquisition. It uses | ||
- | '''Cycle Composer Software'''[http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-hts-pal-and-ctc-htc-pal-software/cycle-composer-remote-control-software-106.php] - this application is run with Standard LEAP macros outside of ''' | + | '''Cycle Composer Software'''[http://www.leaptec.com/products/lc-related-instruments-and-accessories/ctc-hts-pal-and-ctc-htc-pal-software/cycle-composer-remote-control-software-106.php] - this application is run with Standard LEAP macros outside of '''Analyst® Software'''[https://products.appliedbiosystems.com/ab/en/US/adirect/ab?cmd=catNavigate2&catID=600927&tab=DetailInfo] from ''Applied Biosystems Inc® '' |
<br><br> | <br><br> | ||
If the peaks of interest come off together in the middle of the gradient (as shown below) then it is possible to utilize the front cut and back cut time to increase throughput. | If the peaks of interest come off together in the middle of the gradient (as shown below) then it is possible to utilize the front cut and back cut time to increase throughput. |
Revision as of 23:14, 28 April 2009
LC Dual Stream Staggered Injections | |
Application Type | |
SPECIAL | |
Application ID | |
M-003 | |
Description | |
High Throughput LC injections alternating onto two analytical columns. |
Overview
This configuration allows shorter cycle times by limiting the acquisition time to just the peaks of interest and regenerating the alternating column during the acquisition. It uses
Cycle Composer Software[1] - this application is run with Standard LEAP macros outside of Analyst® Software[2] from Applied Biosystems Inc®
If the peaks of interest come off together in the middle of the gradient (as shown below) then it is possible to utilize the front cut and back cut time to increase throughput.
If injections are staggered in the following way, there is a potential for up to 100% gain:
Hardware Requirements
- TrioVS multiple valve system is required in addition to the standard injection valve.
Code | Application | Control SW | Key Hardware | Key Benefits | Restrictions | GLP |
---|---|---|---|---|---|---|
M-003 | Dual Channel Parallel Injection Provides shorter cycle times by limiting the acquisition time to just the peaks of interest and regenerating columns during acquisition | Cycle Composer outside of Analyst | * Trio VS *6 port injection valve *FWS in front of 6 port valve * Gradient Pump * Regeneration pump | * Saving of Front and Back Cut * Provides "Look Ahead" *Provides staggering of injections | * Requires matching sample lists in Cycle Composer and Analyst * Columns may perform slightly differently on alternating samples | No |