There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.<br> Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle. | There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.<br> Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle. | ||
+ | |||
+ | |||
+ | === Contact LEAP === | ||
+ | |||
+ | * [http://www.leaptec.com/ LEAP Technologies, Inc.] | ||
+ | |||
+ | Keyword: Analyst Software, ICC-CE, PAL,TRIO VS, 10 port valve, selector valve, injector valve, gradient pump, regeneration pump,back cut, peaks of interest, GLP |
Revision as of 13:55, 24 February 2009
Overview
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This can be run directly from within Analyst using a custom Cycle (requires ICC-CE). With the addition of a selector valve, the solvent front can be cut to waste.
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.
Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle.