M-002 Dual Channel Serial Injection with front cut - Analyst with ICC-CE

From LEAP

(Difference between revisions)
Jump to: navigation, search
Line 8: Line 8:
=== Overview ===
=== Overview ===
-
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This can be run directly from within '''Analyst''' using a custom Cycle (requires ICC-CE). With the addition of a selector valve, the solvent front can be cut to waste.
+
This configuration allows the front cut of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within '''Analyst''' using a Custom LEAP Cycle (requires ICC-CE software).
-
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.<br> Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle.
+
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.<br> All three valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle.
 +
 
 +
 
 +
[[Image:key_Benefits2.png‎]]
 +
 
 +
 
 +
 
 +
[[Image:Restrictions.png‎]]
 +
 
 +
 
 +
[[Image:Table M-001.png]]
 +
 
 +
 
 +
=== Power Point of Regeneration ===
 +
*[[:Image:Column_Regeneration.pps| Valve Power Point]]
Line 17: Line 31:
* [http://www.leaptec.com/ LEAP Technologies, Inc.]
* [http://www.leaptec.com/ LEAP Technologies, Inc.]
-
Keyword: Analyst Software, ICC-CE, PAL,TRIO VS, 10 port valve, selector valve, injector valve, gradient pump, regeneration pump,front cut, peaks of interest, GLP
+
Keyword: Analyst Software, ICC-CE, PAL, TRIO Valve, 10 port valve, selector valve, injector valve, gradient pump, regeneration pump,back cut, peaks of interest, GLP

Revision as of 18:41, 25 March 2009

Dual Channel Serial Injection with Front Cut
Application Type
  SPECIAL
Application ID
  M-002
Description
  High throughput where one column is regenerating while the other is performing analytical separation. An additional valve allows solvent front to be cut to waste.

Overview

This configuration allows the front cut of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This is run directly from within Analyst using a Custom LEAP Cycle (requires ICC-CE software).

There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.
All three valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle.


Image:key_Benefits2.png‎


Image:Restrictions.png‎


Image:Table M-001.png


Power Point of Regeneration