M-001 Dual Channel Serial Injection - Analyst with ICC-CE
From LEAP
(Difference between revisions)
(New page: {{LSApp |name = LC Column Switching & Regeneration |image = TwoValvePic.png |type = '''SPECIAL''' |id = M-001 |description = LC Column Switching & Re...) |
(→Overview) |
||
Line 22: | Line 22: | ||
[[Image:Restrictions.png]] | [[Image:Restrictions.png]] | ||
+ | |||
+ | |||
+ | Keyword: Analyst Software, ICC-CE, PAL, 10 port valve, gradient pump, regeneration pump,back cut, peaks of interest, GLP |
Revision as of 13:36, 24 February 2009
LC Column Switching & Regeneration | |
Application Type | |
SPECIAL | |
Application ID | |
M-001 | |
Description | |
LC Column Switching & Regeneration. |
Overview
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This can be run directly from within Analyst using a custom Cycle (requires ICC-CE).
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.
Both valves must be controlled by the custom cycle. LEAP can not control the MS Selector valve as part of the cycle.
Keyword: Analyst Software, ICC-CE, PAL, 10 port valve, gradient pump, regeneration pump,back cut, peaks of interest, GLP