LC Dual Stream Staggered Injections
From LEAP
(Difference between revisions)
(New page: {{LSApp |name = LC Dual Stream Staggered Injections |image = Agilent 6890 GC with 5975C MSD.jpg |type = '''SPECIAL''' |id = LC1003 |description = Hig...) |
Current revision (17:07, 4 February 2009) (view source) |
||
(5 intermediate revisions not shown.) | |||
Line 1: | Line 1: | ||
{{LSApp | {{LSApp | ||
|name = LC Dual Stream Staggered Injections | |name = LC Dual Stream Staggered Injections | ||
- | |image = | + | |image = M-Fab_Photo1.png |
|type = '''SPECIAL''' | |type = '''SPECIAL''' | ||
|id = LC1003 | |id = LC1003 | ||
|description = High Throughput LC injections alternating onto two analytical columns. | |description = High Throughput LC injections alternating onto two analytical columns. | ||
}} | }} | ||
+ | |||
+ | === Control Options === | ||
+ | |||
+ | * Analyst Cycles (with ICC-CE) - '''NOT POSSIBLE''' Analyst does not support "Look-Ahead" | ||
+ | |||
+ | * Cycle Composer - Supports this application with custom macros outside of Analyst. | ||
+ | |||
+ | *LEAP Shell - '''M-Fab''' application integrated with Analyst with supported pumps. '''NOT A RELEASED PRODUCT''' (Looking for Beta Sites) | ||
=== Overview === | === Overview === | ||
+ | If the peaks of interest come off together in the middle of the gradient (as shown below) then it is possible to utilize the front cut and back cut time to increase throughput. | ||
+ | |||
+ | [[Image:Front-backcut.png]] | ||
+ | |||
+ | If injections are staggered in the following way, there is a potential for up to 100% gain: | ||
+ | |||
+ | |||
+ | [[Image:Timeline_Dual_Stream_1.png]] | ||
+ | |||
+ | |||
+ | === Hardware Requirements === | ||
+ | * TrioVS multiple valve system is required in addition to the standard injection valve. | ||
+ | |||
+ | |||
- | + | [[Image:Valve_Drawing_Dual_Stream_1.png]] |
Current revision
LC Dual Stream Staggered Injections | |
Application Type | |
SPECIAL | |
Application ID | |
LC1003 | |
Description | |
High Throughput LC injections alternating onto two analytical columns. |
Control Options
- Analyst Cycles (with ICC-CE) - NOT POSSIBLE Analyst does not support "Look-Ahead"
- Cycle Composer - Supports this application with custom macros outside of Analyst.
- LEAP Shell - M-Fab application integrated with Analyst with supported pumps. NOT A RELEASED PRODUCT (Looking for Beta Sites)
Overview
If the peaks of interest come off together in the middle of the gradient (as shown below) then it is possible to utilize the front cut and back cut time to increase throughput.
If injections are staggered in the following way, there is a potential for up to 100% gain:
Hardware Requirements
- TrioVS multiple valve system is required in addition to the standard injection valve.