LC Dual Stream Staggered Injections

From LEAP

(Difference between revisions)
Jump to: navigation, search
(New page: {{LSApp |name = LC Dual Stream Staggered Injections |image = Agilent 6890 GC with 5975C MSD.jpg |type = '''SPECIAL''' |id = LC1003 |description = Hig...)
Current revision (17:07, 4 February 2009) (view source)
 
(5 intermediate revisions not shown.)
Line 1: Line 1:
{{LSApp
{{LSApp
   |name        = LC Dual Stream Staggered Injections
   |name        = LC Dual Stream Staggered Injections
-
   |image      = Agilent 6890 GC with 5975C MSD.jpg
+
   |image      = M-Fab_Photo1.png
   |type        = '''SPECIAL'''
   |type        = '''SPECIAL'''
   |id          = LC1003
   |id          = LC1003
   |description = High Throughput LC injections alternating onto two analytical columns.
   |description = High Throughput LC injections alternating onto two analytical columns.
}}
}}
 +
 +
=== Control Options ===
 +
 +
* Analyst Cycles (with ICC-CE) -  '''NOT POSSIBLE'''  Analyst does not support "Look-Ahead"
 +
 +
* Cycle Composer - Supports this application with custom macros outside of Analyst.
 +
 +
*LEAP Shell - '''M-Fab''' application integrated with Analyst with supported pumps. '''NOT A RELEASED PRODUCT''' (Looking for Beta Sites)
=== Overview ===
=== Overview ===
 +
If the peaks of interest come off together in the middle of the gradient (as shown below) then it is possible to utilize the front cut and back cut time to increase throughput.
 +
 +
[[Image:Front-backcut.png]]
 +
 +
If injections are staggered in the following way, there is a potential for up to 100% gain:
 +
 +
 +
[[Image:Timeline_Dual_Stream_1.png]]
 +
 +
 +
=== Hardware Requirements ===
 +
* TrioVS multiple valve system is required in addition to the standard injection valve.
 +
 +
-
A
+
[[Image:Valve_Drawing_Dual_Stream_1.png]]

Current revision

LC Dual Stream Staggered Injections
Application Type
  SPECIAL
Application ID
  LC1003
Description
  High Throughput LC injections alternating onto two analytical columns.

Control Options

  • Analyst Cycles (with ICC-CE) - NOT POSSIBLE Analyst does not support "Look-Ahead"
  • Cycle Composer - Supports this application with custom macros outside of Analyst.
  • LEAP Shell - M-Fab application integrated with Analyst with supported pumps. NOT A RELEASED PRODUCT (Looking for Beta Sites)

Overview

If the peaks of interest come off together in the middle of the gradient (as shown below) then it is possible to utilize the front cut and back cut time to increase throughput.

Image:Front-backcut.png

If injections are staggered in the following way, there is a potential for up to 100% gain:


Image:Timeline_Dual_Stream_1.png


Hardware Requirements

  • TrioVS multiple valve system is required in addition to the standard injection valve.


Image:Valve_Drawing_Dual_Stream_1.png