=== Overview ===
=== Overview ===
-
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This can be run from Cycle Composer or within Analyst using a custom Cycle (requires ICC-CE).
+
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This can be run directly from within '''Analyst''' using a custom Cycle (requires ICC-CE).
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.  
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.  

Revision as of 15:30, 4 February 2009

LC Column Switching & Regeneration
Application Type
  High Throughput
Application ID
  LC1001
Description
  LC Column Switching & Regeneration.

Overview

This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput. This can be run directly from within Analyst using a custom Cycle (requires ICC-CE).

There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.


Plumbing diagram for each selected column
Plumbing diagram for each selected column