LC Column Switching and Regeneration

From LEAP

(Difference between revisions)
Jump to: navigation, search
(Undo revision 110 by Psmith (Talk))
Line 3: Line 3:
   |image      = TwoValvePic.png
   |image      = TwoValvePic.png
   |type        = High Throughput
   |type        = High Throughput
-
   |id          = LC
+
   |id          = LC1001
-
   |description = LC Column Switching.
+
   |description = LC Column Switching & Regeneration.
}}
}}
=== Overview ===
=== Overview ===
 +
This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput.
 +
There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.

Revision as of 14:56, 4 February 2009

LC Column Switching & Regeneration
Application Type
  High Throughput
Application ID
  LC1001
Description
  LC Column Switching & Regeneration.

Overview

This configuration allows the regeneration of one of the analytical columns while the other one is engaged - giving greater efficiency and higher throughput.

There are 2 pumps needed, one for the gradient run and another (higher flow rate and isocratic) for regeneration. This configuration requires a 6 port valve for sample loading and a 10 port valve for column switching.



Image:Valve_Drawing_ColRegen.png