Sideport Syringe
From LEAP
(Difference between revisions)
(New page: {{LSApp |name = Large Volume Dispensing |image = Pump_with_selector_valve.png |type = '''SPECIAL''' |id = |description = Accurate dispensing of solv...) |
|||
Line 9: | Line 9: | ||
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]] | [[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]] | ||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
=== Solvent Delivery System (SDS) Accessory === | === Solvent Delivery System (SDS) Accessory === |
Revision as of 18:26, 2 June 2009
Large Volume Dispensing | |
Application Type | |
SPECIAL | |
Application ID | |
Description | |
Accurate dispensing of solvents at high flow rates |
Contents |
Solvent Delivery System (SDS) Accessory
Main Features
- 4 Solvents can be dispensed
- 9 Program Steps allow solvents to be added in any order
- Software Interface SDS Launchpad for loading dispense routines
- Save/Load Methods to PC drive
- Easy Verification of Downloaded Method via the “Dashboard” feature in SDS Launchpad
- Contact closure triggers device to run stored dispense routine
- Significantly extends the PAL dispenseing volume range.
- Can be used with any PAL side port syringe or LEAP DISPENSE Tool
- Compatible with all PAL Drivers
Performance Specifications
- Maximum flowrate 6mL/min in increments of 0.1mL/min
- Flowrate accuracy ± 10%
- Maximum dispensed volume 999mL
- Volumetric accuracy ± 5%
- Maximum pressure 50PSIG
- Typical wash time for 2 solvents (1mL each) @ 6mL/min is 45 seconds
Requirements/Limitations
- Lower dispense speeds needed for more viscous solvents, e.g., i-Propanol @ 3-4mL/min
- Use of Polypropylene compatible solvents
Screenshots
Side Port Syringe Accessory
Main Features
- Can be used to dispense liquid prior sample injection
- Combines the "add liquid" and "inject" steps without user intervention or syringe change
- Variety of sizes available: 20µl or 80µl for LC applications and 20µl or 80µl for GC applications
PRINT Function for Page
Contact LEAP
For additional information about this technique please contact LEAP Technologies for detailed information