Monitoring of Micro-dissolution

From LEAP

(Difference between revisions)
Jump to: navigation, search
Current revision (13:31, 8 June 2010) (view source)
(Significant Markets)
 
(4 intermediate revisions not shown.)
Line 6: Line 6:
   |description = Sampling of slow release drugs from micro-dissoultion vessels
   |description = Sampling of slow release drugs from micro-dissoultion vessels
}}
}}
-
 
+
{{logo}}
=== Overview ===
=== Overview ===
Dissolution testing of slow-release devices, for example small, polymer coated springs or coils which can be implanted inside the human body and release their drugs over a period of many months or even years. It is applicable to any customer who is developing or manufacturing these devices. This type of testing is a necessary part of the development and or QC process. The objective is to carefully measure the release rate of the drug from the device in-vitro over a period of time. Results are then correlated with data collected in animals.  
Dissolution testing of slow-release devices, for example small, polymer coated springs or coils which can be implanted inside the human body and release their drugs over a period of many months or even years. It is applicable to any customer who is developing or manufacturing these devices. This type of testing is a necessary part of the development and or QC process. The objective is to carefully measure the release rate of the drug from the device in-vitro over a period of time. Results are then correlated with data collected in animals.  
Line 31: Line 31:
   Image:Dissolution.png | LEAP Dissolution Testing System
   Image:Dissolution.png | LEAP Dissolution Testing System
</gallery>
</gallery>
-
 
+
=== Videos of PAL ===
-
-----
+
<br>
-
 
+
[[Image:Movie Icon.png|40px]]
-
[[Image:LEAP Logo.png|frame|[http://www.leaptec.com leaptec.com]<br />[http://www.leapwiki.com leapwiki.com]]]
+
[http://www.youtube.com/user/LEAPTechnologies#g/u LEAP's PAL Application Videos on YouTube]
 +
<br>
 +
----
 +
{{logo}}
 +
LEAP provides automated workstation instrumentation solutions based on the LEAP CTC PAL X, Y, Z syringe only autosampler robot from LEAP Technologies. This extremely flexible, precise, and adaptable liquid handling robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your UHPLC, LC or GC chromatography.LEAP offers full support and service for the PAL platform in addition to being able to write custom macros, cycles, and scheduling to your applications. Please contact LEAP Technologies on how we can help you get maximized throughput with flexible pipetting automation solutions.
=== Contact LEAP ===
=== Contact LEAP ===
-
 
+
{{contact|topic=LEAP and the PAL Platform}}
-
For additional information about this technique please contact LEAP Technologies for detailed information
+
-
 
+
-
[[image:Contact_Info.png‎|40px]] [http://www.leaptec.com/contact-us.php Contact the LEAP office]
+
-
 
+
-
<br><br>
+
[[Category:LEAP Shell]]
[[Category:LEAP Shell]]
[[Category:Process Monitoring]]
[[Category:Process Monitoring]]
 +
[[Category:Application_Solutions]]

Current revision

Monitoring of Dissolved Product
Application Type
  Process Monitoring
Application ID
  ProcessMonitoring1001
Description
  Sampling of slow release drugs from micro-dissoultion vessels
 ↳ leaptec.com  ↳ leapwiki.com

Contents

Overview

Dissolution testing of slow-release devices, for example small, polymer coated springs or coils which can be implanted inside the human body and release their drugs over a period of many months or even years. It is applicable to any customer who is developing or manufacturing these devices. This type of testing is a necessary part of the development and or QC process. The objective is to carefully measure the release rate of the drug from the device in-vitro over a period of time. Results are then correlated with data collected in animals.

Significant Markets

This application is ideal for those labs who wish to perform dissolution testing in small vessels. Samples or devices are normally held in vials incubated and shaken for prolonged periods and elution volumes removed and tested periodically. Elements of this application could be applied to labs requiring any type of sample treatments at timed intervals. It also involves storing data about each sample in Excel and protecting the devices from contact with the sampling needle

What is a Dissolution Testing experiment?
The testing procedure is quite simple. Vials containing the devices are filled with a buffer solution. They are loaded into a special heated and shaking tray. A timer is started. At specified intervals the buffer is removed, a portion of it saved for later analysis, the rest discarded, and the vials replenished with fresh buffer. Despite its simplicity, this process requires the manual sampling and testing of hundreds of vials at very carefully measured time intervals over a period of days. It is a very tedious operation which requires technicians to be in and out of the lab for hours doing a very mundane job – a classic candidate for automation. It was important that any automation solution did not damage the surface of the devices. There were no existing options to automate this process before the customer approached LEAP.

LEAPS Approach
The PAL in combination with the LEAP Shell makes a very effective and flexible workstation.
We were able to design and provide a solution which met the following challenges from the customer:
1. Exchanging large volumes of liquid from sealed vials. This requires overcoming vacuum/pressure differentials.
2. Protect the slow-release devices from contact and potential damage by the sampling needle. This was achieved using a specially designed vial insert.
3. Shaking and incubating a large number of sample vials (>20). We were able to source a suitable heated shaking table for their vials from a third party vendor.
4. Setting up a sampling schedule in the software which would span over several days if necessary. Only possible with the LEAP Shell.
5. Provide the flexibility to change the sample batch size and sampling intervals.
6. Record the actual sampling times to a data file – compare actual time with expected time and report a difference.

Photos

Videos of PAL


LEAP's PAL Application Videos on YouTube


 ↳ leaptec.com  ↳ leapwiki.com

LEAP provides automated workstation instrumentation solutions based on the LEAP CTC PAL X, Y, Z syringe only autosampler robot from LEAP Technologies. This extremely flexible, precise, and adaptable liquid handling robotic platform is available in a variety of lengths and options depending on the requirements of your sample preparation and injections for your UHPLC, LC or GC chromatography.LEAP offers full support and service for the PAL platform in addition to being able to write custom macros, cycles, and scheduling to your applications. Please contact LEAP Technologies on how we can help you get maximized throughput with flexible pipetting automation solutions.

Contact LEAP

Contact LEAP

For additional information about LEAP and the PAL Platform, please contact LEAP Technologies.

Personal tools